Acquiring and interpreting steady state fluorescence spectra is not always straightforward. Below we offer some advice on commonly encountered problems to help you optimise the operation of your Edinburgh Instruments spectrometer.
At high intensities, PMT detectors are not able to count all photons and the sensitivity of the instrument stops being linear. This is known as detector saturation. Although it may not damage the detector (at least at moderate intensities), it results in spectral distortion and must therefore be avoided. The saturation limit depends on the type of PMT detector. A typical threshold for standard visible PMT’s is ~ 1.5×10⁶ cps.
Some tips on avoiding saturating are listed below: